12 13
FIGURE 2.10B
TIP - If realigning existing
probe bracket, loosen (DO NOT
REMOVE) 4 mounting screws
and compression fitting con-
nections to allow repositioning
of probe bracket steam ports.
2.10 – ALIGNMENT TOOL
(RPI PART #SCT052)
When removing or replacing the probe bracket assembly,
the steam ports must be properly aligned. Use this Align-
ment Tool to easily and accurately set the position of the
steam ports to the cassette.
INSTRUCTIONS
Separate the cassette tray from the cassette lid. Place
the lid on counter top and remove the cassette seal. The
cassette seal must be removed before continuing with
Alignment Tool installation (see page 10 - Figure 2.6).
STEP #1 - Tilt the cassette lid and install Alignment Tool
into steam ports from the inside of the cassette lid as
shown. The square tabs on the tool must engage the
square notches in the cassette lid, and the tool must lay
flat within the cassette seal groove. The Alignment Tool
must be seated correctly to avoid damaging the tray when
FIGURE 2.10A
TIP - Angle cassette lid as shown, using
gravity to aid in tool positioning while
assembling and closing the cassette.
latched and closed. See Figure 2.10A.
(Caution - Ensure that the Alignment Tool is properly in-
stalled before closing the cassette lid or damage can occur.)
STEP #2 - Carefully attach the bottom tray to the cassette
lid and close to properly retain the Alignment Tool. The cas-
sette lid should close normally and retain the Alignment
Tool. DO NOT FORCE IT CLOSED!
STEP #3 - The cassette is now ready to be used to align
probe bracket steam ports.
STEP #4 - If realigning an existing probe bracket, loosen
(DO NOT REMOVE) 4 mounting screws (using 9/64" Hex
Balldriver - RPI Part #RPT836) and compression fitting con-
nections to allow repositioning of probe bracket steam ports.
STEP #5 - Slowly insert the cassette into the cassette bay
(armature). As the cassette approaches the probe bracket
(still loosely installed), position bracket assembly up and
down and sideways until probe bracket steam ports and
cassette engage easily. Once properly inserted, the spring
bar should engage and lock the cassette into position. A
"click" will be heard and felt. See Figure 2.10B.
STEP #6 - Lightly snug the probe bracket mounting screws.
Move the cassette in and out several times to ensure proper
alignment. Leave the cassette in the engaged position. Now
tighten mounting screws to retain bracket position. The
probe bracket assembly is now aligned and ready for use.
STEP #7 -
Remove Alignment Tool. Reinstall or replace Cas-
sette Seal (RPI Part #SCS001 fits STATIM 2000 & RPI Part
#SCS029 fits STATIM 5000) using Liquid Soap (RPI Part
#RPS287) included with the probe brackets. See page 10 -
Figure 2.7. For additional information regarding the care and
maintenance of the cassette seal, see “Installation and
Monthly Maintenence Guide” that is supplied with RPI's Cas-
sette Seals (RPI Part #'s SCS001 & SCS029) or see page 22.
STEP #8 - After successful installation or realignment of
probe bracket, install or realign thermocouple. See RPI Part
#SCT030 Installation Instructions that are available on the
RPI website (www.rpiparts.com).
STEP #9 - Coat entire inside of cassette with Sci-Dry™
(RPI Part #SCA054).
RELATED RPI PARTS:
• Probe Bracket Kit 2000 (RPI Part #SCK038)
• Probe Bracket Kit 5000 (RPI Part #SCK042)
• Probe Bracket 2000 (RPI Part #SCB039)
• Probe Bracket 5000 (RPI Part #SCB043)
Probe Bracket Gasket (RPI Part #SCG010)
• Cassette Seal 2000 (RPI Part #SCS001)
• Cassette Seal 5000 (RPI Part #SCS029)
• Microswitch Kit (RPI Part #SCK007)
Thermocouple (RPI Part #SCT030)
• Liquid Soap (RPI Part #RPS287)
• Hex Balldriver (9/64") (RPI Part #RPT836)
Thermocouple
(RPI Part #SCT030)
Microswitch Kit
(RPI Part #SCK007)
Installed
Grounding Wire
Probe Bracket Kit
2000 (RPI Part #SCK038); 5000 (RPI Part #SCK042)
Probe Bracket
2000 (RPI Part #SCB039); 5000 (RPI Part #SCB043)
Internal View of
Cassette Bay
(Armature)
Spring Bar
Probe Bracket Gasket
(RPI Part #SCG010)
Hex Balldriver (9/64")
(RPI Part #RPT836)
Alignment Tool
(RPI Part #SCT052)
Exhaust Steam Port
Cassette Lid
square notches
Injection Steam Port
Seitenansicht 7
1 2 3 4 5 6 7 8 9 10 11 12 13 ... 28 29

Kommentare zu diesen Handbüchern

Keine Kommentare