6.0 – LCD ERROR CODES
(continued)
STAGE #5 – VENTING
(continued)
LCD MESSAGE
CYCLE FAULT #16
Steam generator/boiler temperature "ran away" (above
171˚C) in the autoclave.
THINGS TO CHECK
• A weak pump due to dirty mesh filters in the pump fit-
tings or dirty Water Filter (RPI Part #SCF034). Check
pump’s operation – see pages 6 and 7, 2.3 - Pump
Tester Bottle (RPI Part #SCK024). If pump passes test,
boiler must be replaced.
Check for continuous power (non-pulsing) to steam gen-
erator/boiler. If continuous power is observed, replace
main PC board.
Steam generator/boiler calibration is needed – see page
40, Figure 9.3 - Steam Generator/Boiler Calibration
Procedure.
LCD MESSAGE
CYCLE FAULT #17 and #18……
Not Currently In Use
LCD MESSAGE
CYCLE FAULT #19
Steam generator/boiler calibration invalid.
THINGS TO CHECK
Steam generator/boiler calibration is needed – see page
40, Figure 9.3 - Steam Generator/Boiler Calibration
Procedure.
If problem repeats, replace the microprocessor/eprom
(matched set) or main PC board, then recalibrate the
steam generator/boiler – see page 40, Figure 9.3 -
Steam Generator/Boiler Calibration Procedure.
LCD MESSAGE
CYCLE FAULT #20
Pump failure.
THINGS TO CHECK
A weak pump due to dirty mesh filters in the pump fit-
tings or dirty W
ater Filter (RPI Part #SCF034). Check
pump’s operation – see pages 6 and 7, 2.3 - Pump
Tester Bottle (RPI Part #SCK024). If pump passes test,
boiler must be replaced.
CYCLE FAULT #20
(continued)
THINGS TO CHECK
(continued)
Steam generator/boiler calibration is needed – see page
40, Figure 9.3 - Steam Generator/Boiler Calibration
Procedure.
LCD MESSAGE
CYCLE FAULT #21 thru #24……
Not Currently In Use
LCD MESSAGE
CYCLE FAULT #25
Unable to power up the steam generator/boiler.
THINGS TO CHECK
Check thermal fuse for continuity.
If needed, replace the
Thermal Fuse (RPI Part #SCF006) – see page 38,
Figure 9.1 - Steam generator/boiler.
A weak pump due to dirty mesh filters in the pump fit-
tings or dirty Water Filter (RPI Part #SCF034). Check
pump’s operation – see pages 6 and 7, 2.3 - Pump
Tester Bottle (RPI Part #SCK024). If pump passes test,
boiler must be replaced.
Steam generator/boiler calibration is needed – see page
40, Figure 9.3 - Steam Generator/Boiler Calibration
Procedure.
LCD MESSAGE
CYCLE FAULT #26
The sterilization stage has failed ~3 minutes after the cas-
sette reaching the sterilization temperature.
F
or 4xx and up software only: “Cycle Interrupted” is dis-
played 2 times, and “Cycle Fault #26” is displayed upon
the 3rd failure.
THINGS TO CHECK
• Check for steam leaks.
• Solenoid valve plunger may be compromised by debris
or may be worn. Check the plunger inside the flute/bon-
net for proper movement. Replace Solenoid-Plunger Kit
(RPI Part #SCK003) if necessary. See page 41, Figure
9.4 - Solenoid Valve.
Steam generator/boiler calibration is needed – see page
40, Figure 9.3 - Steam Generator/Boiler Calibration
Procedure.
6.0 – LCD ERROR CODES
(continued)
STAGE #5 – VENTING
(continued)
LCD MESSAGE
CYCLE FAULT #27
The steam generator/boiler temperature has exceeded
150˚C for 25 seconds (or a time out period).
THINGS TO CHECK
A weak pump due to dirty mesh filters in the pump fit-
tings or dirty Water Filter (RPI Part #SCF034). Check
pump’s operation – see pages 6 and 7, 2.3 - Pump
Tester Bottle (RPI Part #SCK024). If pump passes test,
boiler must be replaced.
Steam generator/boiler calibration is needed – see page
40, Figure 9.3 - Steam Generator/Boiler Calibration
Procedure.
LCD MESSAGE
CYCLE FAULT #72 and 98
Communication error between the microprocessor and the
temperature adapter board.
If your autoclave is equipped with the temperature adapter
board (Alex Upgrade), the following two Cycle Faults may
apply – see page 25, Service Tip #9.
THINGS TO CHECK
Check connection between the “square” microproces-
sor and the main PC board. Also check the connection
between the temperature adapter board and the main
PC board. Ensure there are no bent or shorted pins.
• Validation thermocouple supplied with this “Alex Boiler
Kit” may require calibration, which is not covered in this
booklet. Consult OEM for this calibration procedure.
STAGE #6 – DRYING
The air compressor forces air through the heated steam generator/boiler into the cassette, therefore drying the instruments.
The drying cycle will run for 60 minutes or until interrupted by pressing the Stop button. Keep in mind that when the Stop but-
ton is pressed, the following message is displayed: “Please Wait - Cycle Complete”. Remove cassette only when the message
displays: “Remove Cassette Cycle Complete".
There are no error codes applicable to this stage.