MEDO PUMP
STATIM 2000 (1990 to present )
Air Compressor Kit (RPI Part
#SCK020) - includes Check Valve
Kit (RPI Part #SCK011).
The MEDO Pump draws air
through the Compressor Filter
(RPI Part #SCF002) located in the
front of the compressor. Air is
pushed through the Biological
Filter (RPI Part #SCF032), if the
unit has one. Then air passes
through the check valve into the
steam generator/boiler.
THOMAS PUMP
STATIM 5000 (2003 thru 2011)
The THOMAS Pump draws unfil-
tered air into sterilizer. Air is then
pushed through the Biological
Filter (RPI Part #SCF032) and
through the check valve into the
steam generator/boiler
MEDO PUMP
STATIM 5000 (1995 thru 2003 &
2011 to present)
The MEDO Pump draws air from
the Air Compressor Filter (RPI
Part #SCF031) mounted to the
outside of the rear chassis of the
sterilizer. Air is then pushed
through the Biological Filter (RPI
Part #SCF032) and through the
check valve into the steam gen-
erator/boiler.
Biological Filter
(RPI Part #SCF032)
Biological Filter
(RPI Part #SCF032)
Biological Filter
(RPI Part #SCF032)
If present
UNITS WITHOUT BIOLOGICAL FILTER
UNITS WITH BIOLOGICAL FILTER
Air Compressor Filter
(RPI Part #SCF031)
Compressor Filter
(RPI Part #SCF002)
Input
Input
Input
CAUTION:
NEVER OIL
CAUTION:
NEVER OIL
CAUTION:
NEVER OIL
Black (#9)
To Main PCB J1-9
White (#10)
To Main PCB J1-10
Output: 2.84 PSI
Output: ~ 4.28 PSI
Output: 4.28 PSI
Steam Generator/Boiler Fitting
- To avoid possible valve failure,
debris must be completely re-
moved from the fitting before
installing new Check Valve (RPI
part #SCK011) using supplied
threadlocker on fitting threads.
Steam Generator/Boiler Fitting
- To avoid possible valve failure,
debris must be completely re-
moved from the fitting before
installing new Check Valve (RPI
part #SCK011) using supplied
threadlocker on fitting threads.
Black (#9)
To Main PCB J1-9
White (#10)
To Main PCB J1-10
Black (#9)
To Main PCB J1-9
White (#10)
To Main PCB J1-10